FLOW DETAILS TAB
FLOW RATE:
This is the instantaneous flow rate and is updated approximately every
100 milliseconds.
FILTERED
This is the filtered flow rate and is identical to the LCD display on the
FLOW
flowmeter (if fitted) and the speed of change will depend on the filter
RATE
setting. The filter will also affect the response time for the analogue
outputs and the flow switches.
TOTAL:
The total liquid passed since the last reset.
PIN 6 & 7:
If no selection is made “Not used” will be in the Notes column. See
following text and chart for the display legends.
OR:- When Pin 6 or 7 is set to “Output” and set to “Pulse” this will
display the frequency of the running output and the number of pulses
per litre selected.
OR:- If it is selected to “Output” and set to “Flow switch” the Value
column will show the transistor status either “low” or “high” and the
notes will show the selected switch points.
OR:- If “Output” and “Reverse Flow” are selected the Value column will
show either “low” or “high” depending on forward or reverse flow.
OR:- If “Input” is selected the Value column will show “high” or “low” depending
on the switch condition and “Input Logic level” in the Notes column.
The large display reflects the meter readings showing on the Atrato LCD Display, if fitted
and will mimic that display. The buttons either side have the functions selected on the
‘Configuration’ screen and also work in parallel with the buttons on the Atrato flowmeter
unit itself. The small window below the rate and total display shows various relevant
operational parameters whilst the unit is operating e.g. relay status and frequency
output if these options are selected; and overfill or under fill if in batch mode.