28
VSPE-de-171117
When mounting the transducer avoid forced twisting or violent
opening. This can damage the transducer!
Overpressure in the vacuum system > 1 bar
Accidental or unintended opening of clamp elements under
stress can lead to injuries due to parts flying around!
Unsecured hose connections can release, process media thus
can leak and possibly damage your health.
Overpressure in the vacuum system 1.5 to 4 bar
KF flange connections with elastomer sealings cannot withstand
such pressures. Process media thus can leak and possibly dam-
age your health.
3.3
Electrical Connection
3.3.1
Voltage Supply / RS485 Interface
The electrical connection is to be made by means of suitable cables considering
EMI demands and according to the pin description shown below:
Connector
Pin 1:
Pin 2:
Pin 3:
Pin 4:
Pin 5:
M12, A coded, 5pin, female
RS485 +
Voltage Supply 24 VDC
GND
RS485 -
n.c.
We recommend to have Shield and supply common (Pin 3)
grounded in the supply unit.
Incorrect connection or inadmissible supply voltage can damage