Thornhill Research Inc.
MOVES
®
SLC™ Operator’s Manual
277
Alarm / Message
Priority
Locked /
Latching
Inhibitable
Causes
Action(s)
112.
IP1: High CVP
Medium
CVP measurement is greater
than or equal to the operator
set high CVP alarm limit.
1. Check for related alarms and
follow actions if required.
2. Clinical intervention may be
required.
113.
IP1: Low ICP
High
ICP measurement is less than
or equal to the operator set
low ICP alarm limit.
1. Check for related alarms and
follow actions if required.
2. Clinical intervention may be
required.
114.
IP1: High ICP
Medium
ICP measurement is greater
than or equal to the operator
set high ICP alarm limit.
1. Check for related alarms and
follow actions if required.
2. Clinical intervention may be
required.
115.
IP2: Measurement
error
Low
IP measurement contains
values that are outside of the
measurement range of the
sensor.
1. Fully disconnect then reconnect IP
patient connection and zero IP
transducer as prompted.
2. Replace IP patient connection.
3. Check integrity of interface with
patient.
4. Power cycle MOVES
®
SLC™.
5. Return MOVES
®
SLC™ for
servicing.
116.
IP2 disconnect
Low
Inhibitable
(Can be set
by
System
Administra-
tor.
to not be
inhibitable)
IP patient connection is
disconnected from system
after previously being
connected.
1. Fully disconnect then reconnect IP
patient connection.
2. Replace IP patient connection.
3. Power cycle MOVES
®
SLC™.
4. Return MOVES
®
SLC™ for
servicing.
Summary of Contents for MOVES SLC
Page 2: ...THIS PAGE DELIBERATELY LEFT BLANK...
Page 14: ......
Page 18: ...MOVES SLC Operator s Manual Thornhill Research Inc 4 THIS PAGE DELIBERATELY LEFT BLANK...
Page 33: ...Thornhill Research Inc MOVES SLC Operator s Manual 19 THIS PAGE DELIBERATELY LEFT BLANK...
Page 34: ......
Page 48: ...MOVES SLC Operator s Manual Thornhill Research Inc 34 THIS PAGE DELIBERATELY LEFT BLANK...
Page 72: ......
Page 168: ...MOVES SLC Operator s Manual Thornhill Research Inc 154 THIS PAGE DELIBERATELY LEFT BLANK...
Page 185: ...Thornhill Research Inc MOVES SLC Operator s Manual 171 Figure 11 10 Accessing the Info Screen...
Page 206: ......
Page 243: ...Thornhill Research Inc MOVES SLC Operator s Manual 229 THIS PAGE DELIBERATELY LEFT BLANK...
Page 244: ......
Page 308: ......
Page 352: ......
Page 354: ...MOVES SLC Operator s Manual Thornhill Research Inc 340 THIS PAGE DELIBERATELY LEFT BLANK...
Page 356: ...MOVES SLC Operator s Manual Thornhill Research Inc 342 THIS PAGE DELIBERATELY LEFT BLANK...