MOVES
®
SLC™ Operator’s Manual
Thornhill Research Inc.
274
Alarm / Message
Priority
Locked /
Latching
Inhibitable
Causes
Action(s)
100.
NIBP set to:
Manual update
Low
Locked &
Latching
Inhibitable
NIBP measurement was not
obtained three consecutive
times when set to take
automatic measurements.
1. Attempt another NIBP reading.
2. Ensure NIBP patient arm cuff is
full connected to MOVES
®
SLC™
and has no leaks.
3. Replace NIBP patient arm cuff.
4. Power cycle MOVES
®
SLC™.
5. Return MOVES
®
SLC™ for
servicing.
101.
NIBP fault
Low
NIBP module internal fault or
NIBP module communication
failure with patient monitor.
1. Power cycle MOVES
®
SLC™.
2. Attempt another NIBP reading.
3. Return MOVES
®
SLC™ for
servicing.
102.
NIBP Mode
changed
STAT -> 5 min
Low
Locked &
Latching
NIBP has been set to STAT
mode for 15 minutes and has
automatically changed to take
readings every 5 minutes for
long term monitoring.
1. No action required. NIBP mode
setting can be changed back to
STAT at any time if required.
103.
IP zero required
Low
Inhibitable
(Can be set
by
System
Administra-
tor.
to not be
inhibitable)
At least one invasive channel
has the following set of
conditions:
a) invasive channel is
connected
b) zero is required
c) channel not currently
zeroing
d) zero failed is not
active on the channel
1. Zero all invasive channels which
require zeroing or disconnect IP
cable(s) if not in use.
2. Fully disconnect then reconnect IP
patient connection and zero IP
transducer as prompted.
3. Replace IP patient connection.
4. Check integrity of interface with
patient.
5. Power cycle MOVES
®
SLC™.
6. Return MOVES
®
SLC™ for
servicing.
Summary of Contents for MOVES SLC
Page 2: ...THIS PAGE DELIBERATELY LEFT BLANK...
Page 14: ......
Page 18: ...MOVES SLC Operator s Manual Thornhill Research Inc 4 THIS PAGE DELIBERATELY LEFT BLANK...
Page 33: ...Thornhill Research Inc MOVES SLC Operator s Manual 19 THIS PAGE DELIBERATELY LEFT BLANK...
Page 34: ......
Page 48: ...MOVES SLC Operator s Manual Thornhill Research Inc 34 THIS PAGE DELIBERATELY LEFT BLANK...
Page 72: ......
Page 168: ...MOVES SLC Operator s Manual Thornhill Research Inc 154 THIS PAGE DELIBERATELY LEFT BLANK...
Page 185: ...Thornhill Research Inc MOVES SLC Operator s Manual 171 Figure 11 10 Accessing the Info Screen...
Page 206: ......
Page 243: ...Thornhill Research Inc MOVES SLC Operator s Manual 229 THIS PAGE DELIBERATELY LEFT BLANK...
Page 244: ......
Page 308: ......
Page 352: ......
Page 354: ...MOVES SLC Operator s Manual Thornhill Research Inc 340 THIS PAGE DELIBERATELY LEFT BLANK...
Page 356: ...MOVES SLC Operator s Manual Thornhill Research Inc 342 THIS PAGE DELIBERATELY LEFT BLANK...