Thornhill Research Inc.
MOVES
®
SLC™ Operator’s Manual
177
Label
Items on the Main Screen
Description
RR
Respiratory Rate, calculated from the CO
2
monitor
•
Numeric 0
–99 B/M in steps of 1, or a series of dashes (- - -)
for unknown value.
•
“>99” is shown if 100 B/M or higher.
PIP
Peak Inspiratory (Airway) Pressure
•
Numeric 0
–100 cmH20 in steps of 1, or a series of dashes
(- - -) for unknown / out of range.
•
Only displayed when ventilating, otherwise, or a series of
dashes (- - -) is shown.
PEEP
Positive End Expiratory (Airway) Pressure
•
Same as PIP
FiO
2
Fraction of Inspired Oxygen (O
2
)
•
Numeric 0
–100 % in steps of 1 or (- - -) for unknown value.
•
If >100 and
≤105, shows 100, if >105 shows (- - -).
•
If unavailable due to calibration / warming-
up shows “CAL”.
ConcFlow
Concentrator Flow
•
Volumetric flow of oxygen from the concentrator into the
ventilator or to O2 Supplementation port (depending on
system mode) at a nominal oxygen concentration of 90%.
•
Numeric 0.0 to 4.0 LPM
•
“>4.0 LPM” shown for any value over 4.0 LPM.
Temp1
Patient Temperature 1
•
Numeric 82.4 to 108.0
°
F (or 28 to 42
°
C) in steps of 0.1 or
(- - -) for unknown value.
•
Values outside of human body temperature range are shown
as follows :
o
“<28.0” or “>42.0” (Celsius)
o
“<82.4” or “>108” (Fahrenheit)
Temp2
Patient Temperature 2
•
Same as Temp1.
PetCO
2
Partial Pressure of End-Tidal Carbon
Dioxide
•
Numeric 0
–99 mmHg in steps of 1, or a series of dashes
(- - -
) for unknown value; otherwise, “>99” is shown.
•
If unavailable due to calibration / warming-
up, shows “CAL”.
PiCO
2
Partial Pressure of Inspired Carbon Dioxide
•
Same as PetCO
2
SpO
2
Saturation percentage of oxygen attached
to hemoglobin.
(The amount of oxygen being carried by the
red blood cells in the blood.)
•
Numeric 0
–100% in steps of 1 or a series of dashes
(- - -) for unknown value.
•
If >100, shows “>100“.
•
If unavailable due to initialization
shows “INIT”.
NOTE:
PVI and the following four (4) “Sp” items are only conditionally displayed when the corresponding feature is
unlocked and the appropriate Masimo sensor is present. These five (5) values can only be found on the left of the pleth
real-time graph.
PVI
Pleth Variability Index
(PVI may help clinicians noninvasively
assess fluid status of patients and predict
fluid responsiveness.)
•
Numeric 0
–100% in steps of 1 or a series of dashes (- - -)
for unknown value.
•
If unavailable due to initialization shows “INIT”.
Summary of Contents for MOVES SLC
Page 2: ...THIS PAGE DELIBERATELY LEFT BLANK...
Page 14: ......
Page 18: ...MOVES SLC Operator s Manual Thornhill Research Inc 4 THIS PAGE DELIBERATELY LEFT BLANK...
Page 33: ...Thornhill Research Inc MOVES SLC Operator s Manual 19 THIS PAGE DELIBERATELY LEFT BLANK...
Page 34: ......
Page 48: ...MOVES SLC Operator s Manual Thornhill Research Inc 34 THIS PAGE DELIBERATELY LEFT BLANK...
Page 72: ......
Page 168: ...MOVES SLC Operator s Manual Thornhill Research Inc 154 THIS PAGE DELIBERATELY LEFT BLANK...
Page 185: ...Thornhill Research Inc MOVES SLC Operator s Manual 171 Figure 11 10 Accessing the Info Screen...
Page 206: ......
Page 243: ...Thornhill Research Inc MOVES SLC Operator s Manual 229 THIS PAGE DELIBERATELY LEFT BLANK...
Page 244: ......
Page 308: ......
Page 352: ......
Page 354: ...MOVES SLC Operator s Manual Thornhill Research Inc 340 THIS PAGE DELIBERATELY LEFT BLANK...
Page 356: ...MOVES SLC Operator s Manual Thornhill Research Inc 342 THIS PAGE DELIBERATELY LEFT BLANK...