MOVES
®
SLC™ Operator’s Manual
Thornhill Research Inc.
164
Setup Screen Options
Fields on the Setup Screen
Used by System Mode
Description
Tidal Volume
Ventilate Modes:
IMV, A/C, SIMV when
Control is set to Volume,
and in PSV with Pressure
Support on and Apnea
Backup on and Control is
set to Volume
•
50 to 250 ml in 10 ml intervals
•
250 to 750 ml in 25 ml intervals
•
Default: 500 ml
•
After a change is accepted, the option selected will be
active after the patient’s current breath.
Control Pressure
Ventilate Modes:
IMV, A/C, SIMV when
Control is set to Pressure,
and in PSV with Pressure
Support on and Apnea
Backup on and Control is
set to Pressure
•
Pressure Control (over PEEP)
•
10 to 55 cmH
2
O in intervals of 1 cmH
2
O
•
Default: 20 cmH
2
O
•
After a change is accepted, the option selected will be
active
after the patient’s current breath.
•
NOTE: Control Pressure = PIP (Peak Inspiratory
Pressure)
– PEEP (Positive End Expiratory Pressure)
PEEP
Ventilate Mode
•
Positive End-Expiratory Pressure
•
0-3, 4 to 20 cmH
2
O in intervals of 1 cmH
2
O
•
Default: 0-3 cmH
2
O
•
NOTE: 0-3 cmH
2
O represents the minimum PEEP setting
for the system. When PEEP is set at: 0-3 cmH
2
O, the
system will maintain a slight bias PEEP pressure of 3
cmH
2
O for improved trigger accuracy and system
responsiveness.
•
After a change is accepted, the option selected will be
active after the patient’s current breath.
•
NOTE: When the PEEP value is increased by 10
cmH
2
O
or more some additional breath triggers may be observed.
This includes starting the ventilator at a PEEP setting of
10
cmH
2
O
or more.
Frequency
All Ventilate modes except
PSV
•
Respiratory Rate
•
6 to 40 B/M in steps of 1 B/M
•
Default: 10 B/M
•
Change at start of next inhalation after change confirmed.
I/E ratio
Ventilate Modes:
IMV, A/C, and in PSV with
Pressure Support on and
Apnea Backup on
•
Inspiratory/Expiratory Ratio
•
1:1, 1:1.5, 1:2, 1:2.5, 1:3.
•
Default: 1:2.
•
Change at start of next inhalation after change confirmed.
Summary of Contents for MOVES SLC
Page 2: ...THIS PAGE DELIBERATELY LEFT BLANK...
Page 14: ......
Page 18: ...MOVES SLC Operator s Manual Thornhill Research Inc 4 THIS PAGE DELIBERATELY LEFT BLANK...
Page 33: ...Thornhill Research Inc MOVES SLC Operator s Manual 19 THIS PAGE DELIBERATELY LEFT BLANK...
Page 34: ......
Page 48: ...MOVES SLC Operator s Manual Thornhill Research Inc 34 THIS PAGE DELIBERATELY LEFT BLANK...
Page 72: ......
Page 168: ...MOVES SLC Operator s Manual Thornhill Research Inc 154 THIS PAGE DELIBERATELY LEFT BLANK...
Page 185: ...Thornhill Research Inc MOVES SLC Operator s Manual 171 Figure 11 10 Accessing the Info Screen...
Page 206: ......
Page 243: ...Thornhill Research Inc MOVES SLC Operator s Manual 229 THIS PAGE DELIBERATELY LEFT BLANK...
Page 244: ......
Page 308: ......
Page 352: ......
Page 354: ...MOVES SLC Operator s Manual Thornhill Research Inc 340 THIS PAGE DELIBERATELY LEFT BLANK...
Page 356: ...MOVES SLC Operator s Manual Thornhill Research Inc 342 THIS PAGE DELIBERATELY LEFT BLANK...