260
Alarm / Message
Priority
Locked /
Latching
Inhibitable
Causes
Action(s)
49. PulseOx: PI reading
confidence poor
Low
Pulse oximeter indicating low
confidence in perfusion index
measurement value due to
low signal quality.
Low signal quality may be a
result of excessive motion
relative to perfusion, or sensor
is damaged or not functioning
1. Check and see if blood flow to the
monitoring site is restricted.
2. Ensure proper sensor type and
application.
3. Minimize or eliminate motion at the
monitoring site.
4. Reapply sensor or select a better
perfused site.
5. Replace pulse oximeter patient
sensor.
6. Power cycle MOVES
®
SLC™.
7. Return MOVES
®
SLC™ for
servicing.
50. PulseOx: SpCO
reading confidence
poor
Low
Pulse oximeter indicating low
confidence in SpCO
measurement value due to
low signal quality.
Low signal quality may be a
result of excessive motion
relative to perfusion, or sensor
is damaged or not functioning
1. Check and see if blood flow to the
monitoring site is restricted.
2. Ensure proper sensor type and
application.
3. Minimize or eliminate motion at the
monitoring site.
4. Reapply sensor or select a better
perfused site.
5. Replace pulse oximeter patient
sensor.
6. Power cycle MOVES
®
SLC™.
7. Return MOVES
®
SLC™ for
servicing.
Summary of Contents for MOVES SLC
Page 2: ...THIS PAGE DELIBERATELY LEFT BLANK...
Page 14: ......
Page 18: ...MOVES SLC Operator s Manual Thornhill Research Inc 4 THIS PAGE DELIBERATELY LEFT BLANK...
Page 33: ...Thornhill Research Inc MOVES SLC Operator s Manual 19 THIS PAGE DELIBERATELY LEFT BLANK...
Page 34: ......
Page 48: ...MOVES SLC Operator s Manual Thornhill Research Inc 34 THIS PAGE DELIBERATELY LEFT BLANK...
Page 72: ......
Page 168: ...MOVES SLC Operator s Manual Thornhill Research Inc 154 THIS PAGE DELIBERATELY LEFT BLANK...
Page 185: ...Thornhill Research Inc MOVES SLC Operator s Manual 171 Figure 11 10 Accessing the Info Screen...
Page 206: ......
Page 243: ...Thornhill Research Inc MOVES SLC Operator s Manual 229 THIS PAGE DELIBERATELY LEFT BLANK...
Page 244: ......
Page 308: ......
Page 352: ......
Page 354: ...MOVES SLC Operator s Manual Thornhill Research Inc 340 THIS PAGE DELIBERATELY LEFT BLANK...
Page 356: ...MOVES SLC Operator s Manual Thornhill Research Inc 342 THIS PAGE DELIBERATELY LEFT BLANK...