MOVES
®
SLC™ Operator’s Manual
Thornhill Research Inc.
272
Alarm / Message
Priority
Locked /
Latching
Inhibitable
Causes
Action(s)
91. Above target volume
Medium
Exhaled tidal volume is at
least 15% greater than the
current operator set point and
patient circuit occlusion alarm
is not active.
1. Check for related alarms and
follow actions if required.
2. Clinical intervention may be
required.
92. Below target volume
Medium
Exhaled tidal volume is at
least 15% less than the
current operator set point and
patient circuit occlusion alarm
is not active.
1. Check for related alarms and
follow actions if required.
2. Clinical intervention may be
required.
93. High expired tidal
volume
Medium
Exhaled tidal volume is 30%
greater than the delivered
inspired tidal volume.
1. Check for related alarms and
follow actions if required.
2. Clinical intervention may be
required.
94. NIBP: Low systolic
pressure
High
Inhibitable
Systolic pressure of NIBP
measurement is less than or
equal to the operator set low
systolic BP alarm limit.
1. Check for related alarms and
follow actions if required.
2. Clinical intervention may be
required.
95. NIBP: High systolic
pressure
High, if
systolic
pressure is
greater
than or
equal to
operator
set high
systolic BP
alarm limit
plus
40 mmHg,
Medium
otherwise.
Inhibitable
Systolic pressure of NIBP
measurement is greater than
or equal to the operator set
high systolic BP alarm limit.
1. Check for related alarms and
follow actions if required.
2. Clinical intervention may be
required.
96. NIBP: Low diastolic
pressure
Medium
Inhibitable
Diastolic pressure of ABP
measurement is less than or
equal to the operator set low
diastolic BP alarm limit.
1. Check for related alarms and
follow actions if required.
2. Clinical intervention may be
required.
Summary of Contents for MOVES SLC
Page 2: ...THIS PAGE DELIBERATELY LEFT BLANK...
Page 14: ......
Page 18: ...MOVES SLC Operator s Manual Thornhill Research Inc 4 THIS PAGE DELIBERATELY LEFT BLANK...
Page 33: ...Thornhill Research Inc MOVES SLC Operator s Manual 19 THIS PAGE DELIBERATELY LEFT BLANK...
Page 34: ......
Page 48: ...MOVES SLC Operator s Manual Thornhill Research Inc 34 THIS PAGE DELIBERATELY LEFT BLANK...
Page 72: ......
Page 168: ...MOVES SLC Operator s Manual Thornhill Research Inc 154 THIS PAGE DELIBERATELY LEFT BLANK...
Page 185: ...Thornhill Research Inc MOVES SLC Operator s Manual 171 Figure 11 10 Accessing the Info Screen...
Page 206: ......
Page 243: ...Thornhill Research Inc MOVES SLC Operator s Manual 229 THIS PAGE DELIBERATELY LEFT BLANK...
Page 244: ......
Page 308: ......
Page 352: ......
Page 354: ...MOVES SLC Operator s Manual Thornhill Research Inc 340 THIS PAGE DELIBERATELY LEFT BLANK...