257
Alarm / Message
Priority
Locked /
Latching
Inhibitable
Causes
Action(s)
43. PulseOx: Low SpCO
perfusion
Low
Inhibitable
Pulse oximeter detects low
perfusion related to SpCO
measurement.
1. Check placement of sensor or if it
is too tight. Confirm that the
sensor is securely on the patient.
2. Reapply sensor or select a better
perfused site.
3. See the directions for use provided
with your sensor.
4. Fully disconnect then reconnect
pulse oximeter patient sensor to
system.
5. Replace pulse oximeter patient
sensor.
6. Clinical intervention may be
required.
44. PulseOx: Low SpMet
perfusion
Low
Inhibitable
Pulse oximeter detects low
perfusion related to SpMet
measurement.
1. Check placement of sensor or if it
is too tight. Confirm that the
sensor is securely on the patient.
2. Reapply sensor or select a better
perfused site.
3. See the directions for use provided
with your sensor.
4. Fully disconnect then reconnect
pulse oximeter patient connection
to system.
5. Replace pulse oximeter patient
sensor.
6. Clinical intervention may be
required.
Summary of Contents for MOVES SLC
Page 2: ...THIS PAGE DELIBERATELY LEFT BLANK...
Page 14: ......
Page 18: ...MOVES SLC Operator s Manual Thornhill Research Inc 4 THIS PAGE DELIBERATELY LEFT BLANK...
Page 33: ...Thornhill Research Inc MOVES SLC Operator s Manual 19 THIS PAGE DELIBERATELY LEFT BLANK...
Page 34: ......
Page 48: ...MOVES SLC Operator s Manual Thornhill Research Inc 34 THIS PAGE DELIBERATELY LEFT BLANK...
Page 72: ......
Page 168: ...MOVES SLC Operator s Manual Thornhill Research Inc 154 THIS PAGE DELIBERATELY LEFT BLANK...
Page 185: ...Thornhill Research Inc MOVES SLC Operator s Manual 171 Figure 11 10 Accessing the Info Screen...
Page 206: ......
Page 243: ...Thornhill Research Inc MOVES SLC Operator s Manual 229 THIS PAGE DELIBERATELY LEFT BLANK...
Page 244: ......
Page 308: ......
Page 352: ......
Page 354: ...MOVES SLC Operator s Manual Thornhill Research Inc 340 THIS PAGE DELIBERATELY LEFT BLANK...
Page 356: ...MOVES SLC Operator s Manual Thornhill Research Inc 342 THIS PAGE DELIBERATELY LEFT BLANK...