© 2015 Thorlabs
3 Commands by Subsystem
79
3.19 Unit Subsystem Commands
For each SCPI command default units are defined, where applicable. The
UNIT
subsystem
provides a mechanism to change these default units. The units selected apply to the
designated command parameters for both command and response
Table <%+%>: Unit Commands
Command
Description
UNIT
Path to unit functions
{C|CEL|CELSius
|F|FAR|FAHRenheit|K|KELVin}
Set the temperature unit
Return the temperature unit
3.19.1 Temperature Unit Commands
Syntax
UNIT:TEMPerature {C|CEL|CELSius|F|FAR|FAHRenheit|K|KELVin}
UNIT:TEMPerature?
Description
The
UNIT:TEMPerature
command sets the temperature unit used for further communication
with the instrument.
Example
UNIT:TEMP K
Sets the temperature unit for further communication with the instrument to Kelvin.
Note
This command is valid only for TED4000 and ITC4000 Series instruments.