3 Commands by Subsystem
53
Command
Description
Query AUTO-PID I value
Query AUTO-PID D value
Query AUTO-PID oscillation
period
Transfer AUTO-PID to control
loop
{MIN|MAX|DEF|<seconds>}
Set protection delay
{MIN|MAX|
DEF|<temperature>}
Set temperature window
amplitude
Query protection tripped
3.11.1 TEC Source Function
Syntax
SOURce{[1]|2}:FUNCtion[:MODE] {TEMPerature|CURRent}
SOURce{[1]|2}:FUNCtion[:MODE]?
Description
The
SOURce{[1]|2}:FUNCtion[:MODE]
command sets the instrument's TEC source function
to current source mode or temperature control mode.
Example
SOUR:FUNC TEMP
Sets the TEC source function of a TED4000 Series instrument to temperature source
mode.
Note
This command is valid for TED4000 and ITC4000 Series instruments only. For TED4000 Series
instruments the command suffix is 1 (can be omitted), for ITC4000 Series instruments the suffix
is 2.
56
56
56
56
62
62
62
62
Summary of Contents for 4000 Series
Page 1: ...Series 4000 LDC40xx TED4015 ITC40xx SCPI Programmer s Reference Manual 2015 ...
Page 2: ...Version Date 3 3 08 Apr 2015 Copyright 2015 Thorlabs ...
Page 69: ... 2015 Thorlabs 3 Commands by Subsystem 67 Figure FIGURENO Status Structure ...