3 Commands by Subsystem
33
3.4.8 Protection Queries
Syntax
OUTPut[1]:PROTection:VOLTage:TRIPped?
OUTPut[1]:PROTection:EXTernal:TRIPped?
OUTPut[1]:PROTection:INTernal:TRIPped?
OUTPut[1]:PROTection:INTLock:TRIPped?
OUTPut[1]:PROTection:KEYLock:TRIPped?
OUTPut[1]:PROTection:OTEMp:TRIPped?
Description
The
OUTPut[1]:PROTection:VOLTage:TRIPped?
query command returns 1 if the LD output
protection voltage was reached.
The
OUTPut[1]:PROTection:EXTernal:TRIPped?
query command returns 1 if the LD-
ENABLE input has a TTL low level.
The
OUTPut[1]:PROTection:INTernal:TRIPped?
query command returns 1 if the
temperature window protection is currently active. This command is valid for ITC4000
instruments only.
The
OUTPut[1]:PROTection:INTLock:TRIPped?
query command returns 1 if the interlock
circuit is open.
The
OUTPut[1]:PROTection:KEYLock:TRIPped?
query command returns 1 if the front panel
key switch is in the lock position.
The
OUTPut[1]:PROTection:OTEMp:TRIPped?
query command returns 1 if the instrument's
internal temperature is too high.
Example
OUTP:PROT:KEYL:TRIP?
Queries the front panel key switch.
0
Typical response.
Note
These commands are valid for LDC4000 and ITC4000 only.