© 2015 Thorlabs
3 Commands by Subsystem
23
Description
Configures the instrument to measure the selected value. The
CONFigure
command does not
perform the measurement. Use the
INITiate
or
READ?
command to take the configured
measurement. The configuration query returns the current measurement configuration.
Example
The following program segment configures a TED4000 Series instrument for temperature
measurements and performs one measurement with the
READ?
command.
CONF:TEMP
Configure the instrument for temperature measurement.
READ?
Perform a measurement and send the reading to the output buffer.
2.01
Typical response.
Example
The following program segment configures a TED4000 Series instrument for temperature
measurements and performs one measurement with the
INITiate[:IMMediate]
and
FETCh?
command sequence.
CONF:TEMP
Configure the instrument for temperature measurement.
INIT
Perform the measurement and store reading in memory.
FETC?
Transfer reading from memory to instrument's output buffer.
2.01
Typical response.
Example
CONF?
Query the current measurement configuration.
TEMP
Typical response.
Example
The following program segment configures a LDC4000 Series instrument for thermopile/power
meter
voltage
measurements
and
performs
one
measurement
with
the
INITiate[:IMMediate]
and
FETCh?
command sequence.
CONF:VOLT2
Configure the instrument for thermopile/power meter voltage measurement.
INIT
Perform the measurement and store reading in memory.
FETC?
Transfer reading from memory to instrument's output buffer.
Summary of Contents for 4000 Series
Page 1: ...Series 4000 LDC40xx TED4015 ITC40xx SCPI Programmer s Reference Manual 2015 ...
Page 2: ...Version Date 3 3 08 Apr 2015 Copyright 2015 Thorlabs ...
Page 69: ... 2015 Thorlabs 3 Commands by Subsystem 67 Figure FIGURENO Status Structure ...
Page 92: ... 2015 Thorlabs 90 Series 4000 SCPI Programmers Reference Manual instrument ...