3.2 System Subsystem Commands
Table <%+%>: System Subsystem Commands
Command
Description
[:IMMediate]
Issues an audible signal
:STATE {ON|1|OFF|0}
Activates/deactivates the beeper
:STATe?
Returns the state of the beeper
[:NEXT]?
Returns the latest <error code>, "Message"
{50|60|AUTO}
Sets the line frequency to fixed 50/60Hz or auto
detect
:ACTual?
Returns the used line frequency
Returns the line frequency setting
Returns the level of SCPI standard (1999.0)
3.2.1 Beeper
Syntax
SYSTem:BEEPer[:IMMediate]
SYSTem:BEEPer:STATe {ON|1|OFF|0}
SYSTem:BEEPer:STATe?
Description
The
SYSTem:BEEPer[:IMMediate]
issues an audible signal.
The
SYSTem:BEEPer:STATe
command activates or deactivates the beeper.
Example
SYST:BEEP
The instrument will issue an audible signal.
SYST:BEEP:STAT?
Queries the state of the beeper activation.
1
Response for beeper activated. The response is ASCII
1
if the beeper is activated or an
ASCII
0
if the beeper is deactivated.
3.2.2 Error Query
Syntax
SYSTem:ERRor[:NEXT]?
Description
The
SYSTem:ERRor[:NEXT]?
command queries the instrument's error queue. This is a query-
only command. For a detailed list of possible error codes see chapter
.
22
16
16
17
17
17
81
Summary of Contents for 4000 Series
Page 1: ...Series 4000 LDC40xx TED4015 ITC40xx SCPI Programmer s Reference Manual 2015 ...
Page 2: ...Version Date 3 3 08 Apr 2015 Copyright 2015 Thorlabs ...
Page 69: ... 2015 Thorlabs 3 Commands by Subsystem 67 Figure FIGURENO Status Structure ...
Page 92: ... 2015 Thorlabs 90 Series 4000 SCPI Programmers Reference Manual instrument ...