15. SENSOR CONNECTION:
x = “+” Sensor connected
x = “-“ Sensor not connected
x = “0” Current flow test of sensor off
Sensor connec.:
xxxxx xxxxx
x = “X” Sensor not configured (see 14. sensor configuration)
Sensor 1…10
x = “?” Sensor not testable
This display shows whether a internal sensor is connected. A “-“ means that the datalogger not
detect the sensor, and the “+” stand for vice versa. A “+” means not mandatory that the sensor is
working respectively a good quality of the measurement.
16. SERIAL INTERFACES MODE:
* COM1: Command
COM2: Sonic R
COM3: SIF002
The display shows the settings of the three serial interfaces.
The COM1 interface is fix adjusted to operate the commands.
To get the lowest power consumption it is necessary to switch COM2 and COM3 off.
For more information to change the serial interface mode for COM2 and COM3 ref. to chap.
Serial command: “Cs” (ref. to chap.
17. MEASURING CYCLE / STORING CYCLE:
*Meas.Cyc. 1 sec
Memory C. 1 min
ExtremeC. 1 h
The display shows the adjustment of the measuring cycle (line 1), the mean (line 2) and the
extreme (line 3) value storing cycle. Refer to chap.
for additional explanation.
Setting options
measuring cycle
(applies only for the 10 internal sensors connected directly to the
datalogger):
1, 2, 3, 4, 5, 6, 10, 12, 15, 20, 30 seconds and
1, 2, 3, 4, 5, 6, 10, 12, 15, 20, 30, 60 minutes
Setting options
mean storing cycle
:
1, 2, 3, 4, 5, 6, 10, 12, 15, 20, 30 and 60 minutes
Setting options
extreme storing cycle
:
1, 2, 3, 4, 5, 6, 10, 12, 15, 20, 30 minutes and
1, 2, 3, 4, 6 hours
Serial commands: “MT”, “ST”, “ET” (ref. to chap.
25 - 56