44 - 56
021579/06/09
„DE“<DdMoYyHhMi>
“de” ”<d><m><y><h><m>
Output of stored extreme data from a specific time
Description of parameters: refer to chap.
"EP"
Output of EEPROM-data (parameter memory) for service purposes
"LL"
Logger status:
Output station name, date and time , state of the AD-transducer,
sensor configuration, measuring and storing rates, voltage of the
accumulator, state of power supply, radiation constant, EEPROM-
status, operation mode, contact output-Timer, last DCF-
synchronisation time and appropriate logger time
"MM"
Instant. measuring values: output with sensor identification (multiline)
"mm"
Instantaneous measuring values: output as single-line data record
(sequence as in the display output)
"DD" Output
Logger-date
"DT"<1..31>
Entering day: Setting of day for the logger-clock
(*)
Response: entered day, logger date
"DM"<1..12>
Entering month: setting of month for the logger-clock
(*)
Response: entered month, logger date
"DJ"<0..99>
Entering year: setting of year for logger-clock
(*)
Response: entered year, logger date
"ZZ"
output
logger-time
"ZH"<0..23>
Entering hour: entering of hour for the logger-clock
(*)
Response: entered hour, logger-time
"ZM" <0..59>
Entering minute: entering of minute for the logger-clock
(*)
.
The second is set to zero.
Response: entered minute, logger-time
„ZS“ <0..59>
Entering second: setting of second for the logger-clock.
(*)
Response: entered second, logger time
"XX"
Output of the station name, instrument type, and software-version.
„XXn“<AAAAA>
Entering station name
„LM“<500..900>
Enter of the lower measuring range in hPa of the air pressure sensor
„LN“<500..900>
Enter of the higher measuring range in hPa of the air pressure sensor
“WV”<0..3>
Wind speed sensor type
(0:COMx, 1:Compact, 2:Classic, 3:FirstClass)
“WW”<0,1>
Wind speed sensor current flow test (0:Off)
“WD”<0..2>
Wind direction sensor type (0:COMx, 1:5-/8-Bit, 2:10-Bit)
„NS“<1,2>
Precipitation sensor type resolution (0.1mm / 0.2mm)
“NT“<0,1>
Precipitation sensor current flow test (0:Off)
„LS“<0,1>
Digital input mode: Event(0) or Length(1)
„LT“<0,1>
Digital input logic: active logic low (0) or active logic high (1)
„SK“<n>’ ’<NN.NNNN>
Enter radiation constant n (n=1…5)
Example: STX „SK1 12.42“ ETX (set constant 1 to 12.42)
“KK“<01..10>’ ’<0,1>
Entering sensor configuration of a internal sensor
<01..10> sensor number
<0,1> 0 -> switch off 1 -> switch on
Example: STX “KK02 1“ ETX (switch on sensor 2)
“FF“
Output of all switch-output timers
“FS“<1..6><HH><MM><LL>
Setting of a switch output-timer: <1..6> timer number,
<HH> starting hour, <MM> starting minute, <LL> power-on time in
minutes
Ex.: STX “FS2083005“ ETX (timer 2, start at 08:30 h, time: 5
minutes)