2.1 MOUNTING
The datalogger is designed for wall-mounting or installation on a mast.
For proper protection against lightning strikes we strongly recommend to connect a preferably short
cable (
≤
1m) with a cross section greater than 6mm² between the case (see
Earth) and the mast (if used) with an earth terminal (e.g. grounding bound or main equipotential
bonding conductor). This connection is also necessary when mains (230VAC) is connected to the
datalogger.
It exists two case variants:
•
9.1756.
1
0.x00x (case version A)
•
9.1756.
0
0.x00x (case version B)
For the mast-installation of case version A seven different pipe clips with a diameter from 46 to
137mm are available. For mast-installation with case version B see chap.