CAUTION:
The range of equipment
described here are not to be used in
conjunction with large quantities of
water. The introduction of water and the
resultant vapour could result in
premature component failure.
WARNING:
Do not use flammable
solvents or combustible materials as fire
or explosion may occur. This equipment
uses components which may be a
source of ignition.
FUNCTION
To provide an accurate controlled temperature between a5
C to +60
C.
PRINCIPLE OF OPERATION
The range of Laboratory Incubators are ideally suited for laboratory bench top
applications or floor.
The incubator body is insulated to minimise heat loss. The outer door which is also
insulated closes on a magnetic gasket.
An Inner Clear Door for viewing products under evaluation whilst not disturbing the
conditions, is standard in each model.
Removable chrome plated open wire shelves are supplied with additional shelf
spacing.
The desired operating temperature is set at the temperature controller. A signal from
the temperature sensor located inside at the stainless steel liner wall is monitored at
the controller. The resultant calculated pulsed output is directed to a solid state relay
which switches power to the heating elements which are bonded to the exterior of the
liner.
The warmed walls of the incubator, gently and uniformly heat the air. The amount of
energy is determined by the sensor.
Air movement is either gravity ‘
G
’ convection, Models: TI-150G and TI-250G or Fan
‘
F
’ Forced, Models: TI-150F, TI-250F, TI-350F and TI-500F.
PERFORMANCE
The operating range for this equipment is a5
C to +60
C with a control
accuracy of
0.1
C.
MAXIMUM TEMPERATURE
For this product is 60
C.