background image

P a g e

 | 

 

17/03/11 

 

IMPORTANT INFORMATION 

 

This manual contains operating and safety information. The operator must read and 
understand the contents of this manual prior to using this equipment. Incorrect 
operation may cause harm or damage the severity is classified by the following Alert 
Boxes. Ensure that this manual is saved for future reference: 
 
 

 

 

 

 

 

 

Warning

 alerts apply when there 

is a Possibility of personal injury. 

 
 

 

 

 

 

 

 

 

Caution

 alerts apply when there 

is a Possibility of damage to the 
Equipment . 

 

To maintain safe operation of this equipment be aware of the 

ALERT BOXES

 

located throughout these operating instructions.

 

 
 

CONTENTS 

 

SAFETY INFORMATION 

ALERT BOXES 

INTRODUCTION 

           THE MANUAL 

           UNPACKING 

           LOCATION 

           INTENDED USE 

           FUNCTION 

PRINCIPLE OF OPERATION 

           PERFORMANCE 

           C-TICK 

INSTALLATION 

MAINTENANCE 

TECHNICAL ASSISTANCE 

EXPLANATION OF CONTROLS 

            MAIN SWITCH 

            CONTROLS 

            ELECTRICAL 

OVER TEMPERATURE INDICATION 

OPERATING THE EQUIPMENT 

OPERATING GUIDELINES FOR 3216 
TIMER/PROGRAMMER 

CHANGING SET TEMPERATURE 

OPERATING THE 3216 PROGRAMMER 

9 TO 12 

CALIBRATION  

13 & 14 

WARRANTY POLICY 

15 

 

 

 

 

 

 

 
 
 

Summary of Contents for TI-150F

Page 1: ...Scientific Equipment Pty Ltd T A Thermoline Scientific ABN 80 000 859 129 10 12 Ross Place Wetherill Park N S W 2164 Australia P O Box 6862 Wetherill Park Delivery Centre N S W 1851 Australia Phone 0...

Page 2: ...Possibility of damage to the Equipment To maintain safe operation of this equipment be aware of the ALERT BOXES located throughout these operating instructions CONTENTS SAFETY INFORMATION 2 ALERT BOXE...

Page 3: ...ssemble the components and inspect for damage Retain the packaging materials until the equipment has been thoroughly tested Notify the detail of any damage without delay to your supplier or Thermoline...

Page 4: ...viewing products under evaluation whilst not disturbing the conditions is standard in each model Removable chrome plated open wire shelves are supplied with additional shelf spacing The desired opera...

Page 5: ...han normal levels of cleanliness The external surface powder coated steel and the inner surface stainless steel may be wiped clean using a damp soft cloth CARING FOR STAINLESS STEEL Stainless steel is...

Page 6: ...it breaker continues to trip it may indicate an electrical fault exists Switch Off the power and consult a qualified technician TEMPERATURE CONTROLLER SENSOR This is a Eurotherm 3216 PID control with...

Page 7: ...use flammable solvents in this cabinet as fire or explosion may result This cabinet contains components that may ignite such materials WARNING 2 Do not introduce water into this chamber as the excessi...

Page 8: ...the temperature ramps from one temperature at a set rate to a second temperature and then dwells at that temperature for a set period of time This ramp dwell program has 4 ramp dwell segments CHANGING...

Page 9: ...our TGSP 2 40 C RMP 2 5 C hr Dwell at 40 C for 5 hours DWEL 2 5 Ramp from 40 C to 45 C in one hour TGSP 3 45 C RMP 3 5 C hr Dwell at 45 C for 5 hours DWEL 3 5 Ramp from 45 C to 50 C in one hour TGSP 4...

Page 10: ...ters shown below TSP 1 Target set point 1 RMP 1 Ramp rate 1 Note with RMP 1 OFF the temperature will rise from ambient to 35 C as quickly as possible DWEL 1 Dwell time 1 3 As for segment 1 described a...

Page 11: ...ur then dwell there for 5 hours 5 As for segment 3 described above use the Scroll key to select in turn the tsp 4 parameter Target set point 4 the RMP 4 parameter Ramp rate 4 and the DWEL 4 parameter...

Page 12: ...tat This symbol refers to Timer status reset which means that the program or timer is in the reset mode and is not running The status can be Res Run or Hold leave the program in reset mode until after...

Page 13: ...temperature displayed and temperature measured at the centre of the workspace For this reason the digital temperature controls have a parameter that can be used to correct the temperature displayed ca...

Page 14: ...will show PV OFS Use the up or down keys to enter an offset of 1 0 C Use the Home key to return to the original display You will now notice that the 1 0 C calibration has been added to the incubator t...

Page 15: ...warranty service is required Thermoline Scientific will endeavour to have an authorised service organisation carry out the work if there is an available agent however if it is decided to return the e...

Page 16: ...ly by Any defect in material or workmanship of or any other defect whatsoever in or unsuitability for any purpose of the goods or any part of the goods or default or negligence on the part of Thermoli...

Reviews: