P a g e
|
11
17/03/11
Note that in this instance “
RMP.2
” is set at 5.0ºC/hour so that the program will
ramp from 35ºC to 40ºC in 1 hour, then dwell there for 5 hours.
4. As for segment 2 described abov
e use the “Scroll” key to select in turn the
“
tsp.3
” parameter (Target set point 3), the “
RMP.3
” parameter (Ramp rate 3),
and the “
DWEL.3
” parameter (Dwell time 3). Use the raise or lower keys to set
the parameters shown below.
Note that in t
his instance “
RMP.3
” is set at 5.0ºC/hour so that the program will
ramp from 40ºC to 45ºC in 1 hour, then dwell there for 5 hours.
5.
As for segment 3 described above use the “Scroll” key to select in turn the
“
tsp.4
” parameter (Target set point 4), the “
RMP.4
” parameter (Ramp rate 4),
and the “
DWEL.4
” parameter (Dwell time 4). Use the raise or lower keys to set
the parameters shown below.
Note that in this instance “
RMP.4
” is set at 5.0ºC/hour so that the program will
ramp from 45ºC to 50ºC in 1 hour, then dwell there for 5 hours.
6.
Use the “Scroll” key to select the last parameter, which is “P.CYCL” or
program cycle. This parameter determines how many times the program will
be repeated. If this is left at “1”, there will be no “repeats” and the temperature
will return to the original set temperature. Unless the incubator has a cooling