Optical Alignment
NOTE:
Check the alignment by loosening the arm in the zero order grating Holder tool
(compress the spring on the zero order grating Holder until the pin can be inserted into
the locking hole, and the tool is locked in the ‘open’ position) and moving the grating arm
to the right. A spectrum will be reflected from the grating onto the entrance slit. Ensure
that this spectrum is centered vertically on the slit. If this requires adjustment of the
mirror, then retighten the zero order grating Holder tool, readjust vertically, and ensure
that the light beam is still centered on the target.
6.
Remove the Mirror B2 Target.
Align Monochromator Mirror B2.
1.
Place the Mirror B2 Target onto the front of the Mirror C.
2.
Align the B2 Monochromator mirror horizontally and vertically until the beam is centered on the Exit
Slit (refer to "Mirror Adjustments") and on the target on Mirror C then remove the target from
Mirror C.
Align Sample Mirror C
1.
Orient the Sample Target in position on the Cell Holder Platform so that the rectangular target is
facing the light beam. Secure the Sample Target with its captive screw. Install the Platform with the
Sample Target into the sample compartment.
2.
Align the Sample Mirror C horizontally and vertically until the beam is centered on the Sample
Target (refer to "Mirror Adjustments").
3.
When the alignment has been completed, remove the Cell Holder Platform. Remove the Sample
Target from the Cell Holder Platform and reinstall the previously installed cell holder onto the
Platform.
4.
Compress the spring on the zero order grating holder until the pin can be inserted into the locking
hole, and the tool is locked in the ‘open’ position.
5.
Loosen and remove the screw securing the zero order grating Holder tool.
6.
Remove the zero order grating Holder tool.
7.
Reinstall the Limit Switch, the filter wheel, the Monochromator cover, the Beam Splitter and turret
cover.
Note:
After any optical change, alignment, or optical component replacement, the instrument
must be calibrated with the proper standards using the Service Diagnostic. Instructions
are included with the software.
6-6
Summary of Contents for Spectronic BioMate 3
Page 1: ......
Page 6: ...This page intentionally left blank vi ...
Page 13: ...Optical Diagram Figure 1 1 Optical Diagram Shown are both UV Vis and Vis Optics 1 5 ...
Page 14: ...Specifications This page intentionally left blank 1 6 ...
Page 22: ...Troubleshooting Guide This page intentionally left blank 3 6 ...
Page 36: ...Routine Maintenance This page intentionally left blank 4 8 ...
Page 72: ......
Page 78: ...Drawings This page intentionally left blank 8 2 ...
Page 79: ...Drawings 8 3 Figure 8 1 Spectronic GENESYS 10 Wiring Diagram ...
Page 80: ...wings 8 4 This page intentionally left blank Dra ...
Page 81: ...Drawings 8 5 Figure 8 2 Keypad Schematic ...
Page 82: ...wings 8 6 This page intentionally left blank Dra ...
Page 84: ...wings 8 8 This page intentionally left blank Dra ...
Page 86: ...Drawings This page intentionally left blank 8 10 ...
Page 88: ...Drawings This page intentionally left blank 8 12 ...
Page 90: ...Drawings This page intentionally left blank 8 14 ...
Page 92: ...Drawings This page intentionally left blank 8 16 ...
Page 94: ...Drawings This page intentionally left blank 8 18 ...
Page 96: ...Drawings This page intentionally left blank 8 20 ...
Page 98: ...Drawings This page intentionally left blank 8 22 ...
Page 100: ...Drawings This page intentionally left blank 8 24 ...
Page 102: ...Drawings This page intentionally left blank 8 26 ...
Page 104: ...Drawings This page intentionally left blank 8 28 ...
Page 106: ...Drawings This page intentionally left blank 8 30 ...
Page 108: ...Drawings This page intentionally left blank 8 32 ...
Page 110: ...Drawings This page intentionally left blank 8 34 ...
Page 112: ...Drawings This page intentionally left blank 8 36 ...
Page 114: ...Drawings This page intentionally left blank 8 38 ...
Page 116: ...Drawings This page intentionally left blank 8 40 ...
Page 118: ...Drawings This page intentionally left blank 8 42 ...
Page 120: ...Drawings This page intentionally left blank 8 44 ...
Page 122: ...Drawings This page intentionally left blank 8 46 ...
Page 124: ...Drawings This page intentionally left blank 8 48 ...
Page 126: ...Drawings This page intentionally left blank 8 50 ...
Page 128: ...Drawings This page intentionally left blank 8 52 ...
Page 130: ...Drawings This page intentionally left blank 8 54 ...
Page 132: ...Drawings This page intentionally left blank 8 56 ...
Page 134: ...Drawings This page intentionally left blank 8 58 ...
Page 136: ...Drawings This page intentionally left blank 8 60 ...
Page 138: ...Drawings This page intentionally left blank 8 62 ...
Page 140: ...Drawings This page intentionally left blank 8 64 ...
Page 142: ...Drawings This page intentionally left blank 8 66 ...
Page 144: ...Drawings This page intentionally left blank 8 68 ...
Page 146: ...Drawings This page intentionally left blank 8 70 ...
Page 148: ...Drawings This page intentionally left blank 8 72 ...
Page 150: ...Drawings This page intentionally left blank 8 74 ...
Page 152: ...Drawings This page intentionally left blank 8 76 ...
Page 166: ...Service Diagnostic Software This page intentionally left blank 9 14 ...
Page 170: ...Technical Support Bulletins This page intentionally left blank 10 4 ...