![Thermo Spectronic BioMate 3 Service Manual Download Page 157](http://html.mh-extra.com/html/thermo/spectronic-biomate-3/spectronic-biomate-3_service-manual_1104233157.webp)
Service Diagnostic Software
Reset Spectronic GENESYS 10
1.
Performs a full power-up reboot sequence.
Query Instrument
1.
Updates the display with the current status of the instrument.
Update Settings
1.
Updates the instrument with the displayed settings.
Calibrate
1.
Performs a complete wavelength calibration of the instrument.
2.
Requires the use of a Didymium Filter.
3.
Must be performed whenever the ROM is changed, or Flashed on the Main board.
4.
Must be performed whenever anything (except the lamp) in the optical path is adjusted, cleaned,
repaired or replaced.
9-5
Summary of Contents for Spectronic BioMate 3
Page 1: ......
Page 6: ...This page intentionally left blank vi ...
Page 13: ...Optical Diagram Figure 1 1 Optical Diagram Shown are both UV Vis and Vis Optics 1 5 ...
Page 14: ...Specifications This page intentionally left blank 1 6 ...
Page 22: ...Troubleshooting Guide This page intentionally left blank 3 6 ...
Page 36: ...Routine Maintenance This page intentionally left blank 4 8 ...
Page 72: ......
Page 78: ...Drawings This page intentionally left blank 8 2 ...
Page 79: ...Drawings 8 3 Figure 8 1 Spectronic GENESYS 10 Wiring Diagram ...
Page 80: ...wings 8 4 This page intentionally left blank Dra ...
Page 81: ...Drawings 8 5 Figure 8 2 Keypad Schematic ...
Page 82: ...wings 8 6 This page intentionally left blank Dra ...
Page 84: ...wings 8 8 This page intentionally left blank Dra ...
Page 86: ...Drawings This page intentionally left blank 8 10 ...
Page 88: ...Drawings This page intentionally left blank 8 12 ...
Page 90: ...Drawings This page intentionally left blank 8 14 ...
Page 92: ...Drawings This page intentionally left blank 8 16 ...
Page 94: ...Drawings This page intentionally left blank 8 18 ...
Page 96: ...Drawings This page intentionally left blank 8 20 ...
Page 98: ...Drawings This page intentionally left blank 8 22 ...
Page 100: ...Drawings This page intentionally left blank 8 24 ...
Page 102: ...Drawings This page intentionally left blank 8 26 ...
Page 104: ...Drawings This page intentionally left blank 8 28 ...
Page 106: ...Drawings This page intentionally left blank 8 30 ...
Page 108: ...Drawings This page intentionally left blank 8 32 ...
Page 110: ...Drawings This page intentionally left blank 8 34 ...
Page 112: ...Drawings This page intentionally left blank 8 36 ...
Page 114: ...Drawings This page intentionally left blank 8 38 ...
Page 116: ...Drawings This page intentionally left blank 8 40 ...
Page 118: ...Drawings This page intentionally left blank 8 42 ...
Page 120: ...Drawings This page intentionally left blank 8 44 ...
Page 122: ...Drawings This page intentionally left blank 8 46 ...
Page 124: ...Drawings This page intentionally left blank 8 48 ...
Page 126: ...Drawings This page intentionally left blank 8 50 ...
Page 128: ...Drawings This page intentionally left blank 8 52 ...
Page 130: ...Drawings This page intentionally left blank 8 54 ...
Page 132: ...Drawings This page intentionally left blank 8 56 ...
Page 134: ...Drawings This page intentionally left blank 8 58 ...
Page 136: ...Drawings This page intentionally left blank 8 60 ...
Page 138: ...Drawings This page intentionally left blank 8 62 ...
Page 140: ...Drawings This page intentionally left blank 8 64 ...
Page 142: ...Drawings This page intentionally left blank 8 66 ...
Page 144: ...Drawings This page intentionally left blank 8 68 ...
Page 146: ...Drawings This page intentionally left blank 8 70 ...
Page 148: ...Drawings This page intentionally left blank 8 72 ...
Page 150: ...Drawings This page intentionally left blank 8 74 ...
Page 152: ...Drawings This page intentionally left blank 8 76 ...
Page 166: ...Service Diagnostic Software This page intentionally left blank 9 14 ...
Page 170: ...Technical Support Bulletins This page intentionally left blank 10 4 ...