8 Program end and removal
03/2007
36
DA0621_a/
8 Program end and removal
DANGER
Do not open the sterilizing chamber (8) until removal is
signalled.
Watch for hot steam and hot water!
Scalding hazard!
Follow the specified operating steps to the letter.
DANGER
When pressure-sealed flasks or bags are
accidentally
sterilized:
Watch for broken glass and splitter!
Watch for hot water spray!
Allow at least 4 hours to elapse at the end of sterilization.
Always wear protective clothing, heat-insulated gloves and eye
protectors.
Do not remove flasks or bags until the reference temperature
has dropped below 70 °C.
The end of the program is signalled both acoustically and visually: a buzzer will
sound, and the green START/STOP key (19) will flash. This indicates that all
phases of the sterilization cycle have been completed. The sterilization phase
display (24) shows the indicator lights for all completed phases.
DANGER
You must wear the following protective items when opening
the sterilizing chamber (8) and removing the sterilized articles:
y
Apron,
y
Eye protectors,
y
Heat-insulated gloves.
Watch for hot surfaces and hot steam!
CAUTION
Before opening, check to see if the indicated pressure has
dropped below the set removal pressure (factory setting: 110
kPa).
♦
Press the green START/STOP key (19).
The acoustic signal will be switched off. The door (2) is released. The pressure in
the sterilizing chamber (8) will adjust automatically to the outside pressure
(pressure compensation).
The displays of the control panel (5) will turn off automatically after 10 minutes.