12 Troubleshooting
03/2007
56
DA0621_a/
Message in the
cycle display
Meaning Remedy
Error 17
Overheating
Heating system does not turn off.
Heating system relay defective or
program control error.
Contact after-sales
service.
Error 20
Microheating
Microheating system (heatable
temperature sensor) cannot regulate
water feed to jacket.
Microheating system defective.
Contact after-sales
service.
Error 22
No heater
Heating coils are defective, or
the heating system control is
defective.
Contact after-sales
service.
Error 23
Partial heater
One heating coil is defective.
Contact after-sales
service.
Err1
in the Chamber
field:
in the Media field:
Sensor broken
Chamber temperature sensor is
affected.
Flexible temperature sensor
(reference temperature sensor) is
affected
12
.
Have technician
replace sensor or
contact after-sales
service.
Err2
in the Chamber
field:
in the Media field:
Sensor shorted
Chamber temperature sensor
Flexible temperature sensor
12)2)2)
(reference temperature sensor).
Have technician
replace sensor or
contact after-sales
service.
Err3
in the Chamber
field:
in the Jacket field:
below the Jacket
field:
Pressure sensor defective
Pressure sensor in sterilizing
chamber
Pressure sensor in jacket
12)
Pressure sensor in seal
12)
Contact after-sales
service.
Table 5: Error codes and their meanings
12)
If present