6 Programs
03/2007
29
DA0621_a/
6.3 Setting the sterilizing time
The selected sterilizing time is shown on the cycle display (16).
Setting range: up to 121 °C 1-40 minutes
7))
from 122 °C 1-30 minutes
7)
We recommend the following sterilizing times:
20 min for sterilization at 121 °C (liquids: solutions and culture media; solids:
glassware and instruments);
20 min for sterilization at 134 °C (destructive sterilization: waste).
Disinfection or steaming times will be governed by the type of articles you want to
disinfect or steam.
The feed water supply is limited; it will determine the maximum disinfection or
steaming times.
Program
Maximum disinfection or steaming times
Liquids / Liquids RM
40 minutes
7)
Table 3: Maximum disinfection or steaming times
You can set the sterilizing time on the control panel (5):
♦
Press the selected program key (18) repeatedly until the sterilizing
temperature appears in the cycle display (Figure 8).
Cycle
Steril. Time
20min
Figure 8: Sterilizing time display
♦
Increase or decrease the sterilizing time using the
▲
key (14) or the
▼
key (15).
The sterilizing time will increase or decrease in one-minute increments with
each press of the
▲
key (14) or the
▼
key (15).
When decreasing sterilizing time, the display will change from the minimum
selectable to the maximum selectable sterilizing times (30/40 min, 999 min
or
∞
). This makes it easy to select long sterilizing times.
6.4 Setting the multiple sterilization number
The multiple sterilization program is a special program. The selected multiple
sterilization number is shown on the cycle display (16).
Setting range: from 1 to 99 sterilization cycles.
7)
)
To facilitate longer sterilizing times, the steam sterilizer can be equipped with the water post-feed
expansion option. In this manner, it is possible to implement sterilizing and steaming times of up to
999 minutes (
∞
for the climatic chamber design). The feed water tank level must be checked at regular
intervals.