6 Programs
The program keys are factory preset for complete program cycles. This means that
in addition to the process itself, the sterilizing time and temperature and certain
options for a sterilizing article are predefined. If you frequently sterilize the same
articles, this will be very comfortable.
Program Sterilization Cooling
Base model:
Instruments
121 °C / 20 min
Fast steam release
Destruction
(Waste)
134 °C / 15 min
Cyclical steam release
Liquids
121 °C / 20 min
Cyclical steam release up
to 100 °C; program end at
90 °C
reference
temperature
With Type RM water recooling system option only:
Liquids RM
121 °C / 20 min
Water jacket cooling,
program end at 90 °C
reference temperature
With media temperature control expansion option only:
Liquids /
Liquids RM /
Liquids RO /
Liquids RG
121 °C / 20 min
Sterilizing time
actuated via media
temperature
Depending on program
(see above), program end
at 90 °C
4)
media
temperature
With Type FA exhaust air filtration (
) only:
All programs
Deaeration through
sterile air filter
Cyclical steam release via
sterile air filter
Table 2: Sterilizing programs
When you select a program, an indicator light in the corner of the selected program
key (18) is lit. Sterilizing temperature, multiple sterilization number (special
program), sterilizing time and program name appear in the cycle display (Figure 6).
Figure 6: Cycle display
Cycle
121 C
20min
Instruments
°