2. Turn the overtemp control knob clockwise at least 2°. The alarm should be silenced
and the overtemp indicator light should go out. The overtemp safety thermostat is
now set a few degrees above the control temperature setpoint. When the chamber
temperature rises to the overtemp control point, the alarm system will activate, power
to the heaters will shut off, and the chamber temperature will be maintained at the
overtemp control point.
Set the Undertemp Safety Thermostat
Allow the chamber temperature and humidity to stabilize, then set the undertemp safety
thermostat as follows:
1. Turn the undertemp control knob slowly clockwise until the audible alarm sounds and
the undertemp indicator lights.
2. Turn the undertemp control knob counterclockwise at least 2° on the scale. The
alarm will silence and the undertemp indicator light goes out.
The undertemp safety thermostat is now set a few degrees below the control temperature
setpoint. When the chamber temperature drops to the undertemp control point, the alarm
system activates, power to the compressor shuts off and the chamber temperature is
maintained at the undertemp control point.
When an undertemp condition occurs, the cause must be determined and corrected
before normal operation under the main temperature controller can be resumed.
NOTE
When an overtemp condition occurs, the cause must be determined and
corrected before normal operation under the main temperature controller can
be resumed.
NOTE
When the chamber temperature control setpoint is changed, the overtemp
safety thermostat must be reset to accommodate the change.
NOTE
The overtemp control is not directly calibrated. The numbers on the dial are for
reference only.
NOTE
Operation of light module
Before operating the environmental chamber with light modules burn-in the
light module for 100 h by creating a program with the Operation Time Switch
Light Modules, with the channels A, B and C ON and running time 100 h.
Summary of Contents for Thermo Scientific 3906
Page 6: ...Table of Contents A 4 Model 3900 Series Thermo Scientific ...
Page 12: ... Safety Notes 1 6 Model 3900 Series Thermo Scientific Chapter 1 ...
Page 22: ... Installation 4 6 Model 3900 Series Thermo Scientific Chapter 4 ...
Page 42: ... Product Description 5 20 Model 3900 Series Thermo Scientific Chapter 5 ...
Page 46: ... Operation 7 2 Model 3900 Series Thermo Scientific Chapter 7 ...
Page 76: ... Disposal 11 2 Model 3900 Series Thermo Scientific Chapter 11 ...
Page 84: ... Specifications 12 8 Model 3900 Series Thermo Scientific Chapter 12 ...
Page 94: ... Service 13 10 Model 3900 Series Thermo Scientific Chapter 13 ...
Page 95: ...Thermo Scientific Model 3900 Series 13 11 Service Chapter 13 ...
Page 99: ...Thermo Scientific Model 3900 Series 13 15 Service Chapter 13 ...
Page 100: ... Service 13 16 Model 3900 Series Thermo Scientific Chapter 13 ...
Page 101: ...Thermo Scientific Model 3900 Series 13 17 Service Chapter 13 ...
Page 102: ... Service 13 18 Model 3900 Series Thermo Scientific Chapter 13 ...