Thermo Scientific
Model 3900 Series |
7-1
Operation |
Chapter 7
Operation
Preparing the Environmental Chamber
The environmental chamber must not be released for operation before all major start-up
activities have been completed (see “Installation” on page 4-1).
Device Check
Prior to starting operation, the following environmental chamber components must be
checked for their correct function:
The door seal in the front frame and solid door must not be damaged.
The glass door must not be damaged.
The shelving components must be installed safely.
Unused sockets of the light modules must be covered with a plug.
Preliminary Cleaning and Disinfecting
Disinfect all interior surfaces with a general-use laboratory disinfectant, such as
quaternary ammonium. Rinse thoroughly with sterile distilled water, then 70% alcohol.
Dry with a clean cloth as needed.
Disinfect the shelf channels and shelves, then rinse with distilled water before installing.
Before using any cleaning or decontamination method except those
recommended by the manufacturer, users should check with the manufacturer
that the proposed method will not damage the equipment.
Summary of Contents for Thermo Scientific 3906
Page 6: ...Table of Contents A 4 Model 3900 Series Thermo Scientific ...
Page 12: ... Safety Notes 1 6 Model 3900 Series Thermo Scientific Chapter 1 ...
Page 22: ... Installation 4 6 Model 3900 Series Thermo Scientific Chapter 4 ...
Page 42: ... Product Description 5 20 Model 3900 Series Thermo Scientific Chapter 5 ...
Page 46: ... Operation 7 2 Model 3900 Series Thermo Scientific Chapter 7 ...
Page 76: ... Disposal 11 2 Model 3900 Series Thermo Scientific Chapter 11 ...
Page 84: ... Specifications 12 8 Model 3900 Series Thermo Scientific Chapter 12 ...
Page 94: ... Service 13 10 Model 3900 Series Thermo Scientific Chapter 13 ...
Page 95: ...Thermo Scientific Model 3900 Series 13 11 Service Chapter 13 ...
Page 99: ...Thermo Scientific Model 3900 Series 13 15 Service Chapter 13 ...
Page 100: ... Service 13 16 Model 3900 Series Thermo Scientific Chapter 13 ...
Page 101: ...Thermo Scientific Model 3900 Series 13 17 Service Chapter 13 ...
Page 102: ... Service 13 18 Model 3900 Series Thermo Scientific Chapter 13 ...