6-1
Start-Up |
Chapter 6
Start-Up
When the humidification system is operational, the environmental chamber may be
started. Preset the controls as follows:
Set the Overtemp Safety Thermostat
For best overall performance of the environmental chamber, the refrigeration switch
should be turned On for most applications. When running Low or No humidity at high
temperatures, the refrigeration switch may be turned Off.
Allow the chamber temperature and humidity to stabilize, then set the overtemp safety
thermostat as follows:
1. Turn the overtemp control knob slowly counterclockwise until the audible alarm
sounds and the overtemp indicator lights.
Parameter
Unit
Overtemp Safety Thermostat
Fully Clockwise
Undertemp Safety Thermostat
Fully Counterclockwise
Main Power Switch
ON
Humidity Controller
Desired Setpoint
Temperature Controller
Desired Setpoint
Door Heater
40% (factory set)
NOTE
When operating the environmental chamber with a light module, always
switch on the cooling
Observe temperature display on the device
The defrost switch must be set to “Auto” when the temperature setpoint is 10°C
or below.
Summary of Contents for Thermo Scientific 3906
Page 6: ...Table of Contents A 4 Model 3900 Series Thermo Scientific ...
Page 12: ... Safety Notes 1 6 Model 3900 Series Thermo Scientific Chapter 1 ...
Page 22: ... Installation 4 6 Model 3900 Series Thermo Scientific Chapter 4 ...
Page 42: ... Product Description 5 20 Model 3900 Series Thermo Scientific Chapter 5 ...
Page 46: ... Operation 7 2 Model 3900 Series Thermo Scientific Chapter 7 ...
Page 76: ... Disposal 11 2 Model 3900 Series Thermo Scientific Chapter 11 ...
Page 84: ... Specifications 12 8 Model 3900 Series Thermo Scientific Chapter 12 ...
Page 94: ... Service 13 10 Model 3900 Series Thermo Scientific Chapter 13 ...
Page 95: ...Thermo Scientific Model 3900 Series 13 11 Service Chapter 13 ...
Page 99: ...Thermo Scientific Model 3900 Series 13 15 Service Chapter 13 ...
Page 100: ... Service 13 16 Model 3900 Series Thermo Scientific Chapter 13 ...
Page 101: ...Thermo Scientific Model 3900 Series 13 17 Service Chapter 13 ...
Page 102: ... Service 13 18 Model 3900 Series Thermo Scientific Chapter 13 ...