Chapter 5
Risk of UV exposure
Use the light module only for its intended application. Don’t use ICH modules for
plant growth or animal hatching.
UV-A and blue light radiation
During inspection or replacement of samples, wear protective goggles with
orange lenses.
Cooling of the environmental chamber must always be in operation, when using
a light module.
If cooling is not in operation, the chamber temperature will rise up to approx.
95°C.
NOTE
The light performance of the lamps has its optimum at a working temperature
of 25°C. At higher or lower temperatures inside the chamber, the light current
decreases.
Summary of Contents for Thermo Scientific 3906
Page 6: ...Table of Contents A 4 Model 3900 Series Thermo Scientific ...
Page 12: ... Safety Notes 1 6 Model 3900 Series Thermo Scientific Chapter 1 ...
Page 22: ... Installation 4 6 Model 3900 Series Thermo Scientific Chapter 4 ...
Page 42: ... Product Description 5 20 Model 3900 Series Thermo Scientific Chapter 5 ...
Page 46: ... Operation 7 2 Model 3900 Series Thermo Scientific Chapter 7 ...
Page 76: ... Disposal 11 2 Model 3900 Series Thermo Scientific Chapter 11 ...
Page 84: ... Specifications 12 8 Model 3900 Series Thermo Scientific Chapter 12 ...
Page 94: ... Service 13 10 Model 3900 Series Thermo Scientific Chapter 13 ...
Page 95: ...Thermo Scientific Model 3900 Series 13 11 Service Chapter 13 ...
Page 99: ...Thermo Scientific Model 3900 Series 13 15 Service Chapter 13 ...
Page 100: ... Service 13 16 Model 3900 Series Thermo Scientific Chapter 13 ...
Page 101: ...Thermo Scientific Model 3900 Series 13 17 Service Chapter 13 ...
Page 102: ... Service 13 18 Model 3900 Series Thermo Scientific Chapter 13 ...