87
Thermo Fisher Scientific
User Guide STP420D/STP420ES
routine operations
Editing and Creating Protocols
Note!
Inserting steps: ‘Insert step’ puts the new step above the highlighted step.
15. Select
OK
to save the protocol.
The system checks the validity of the protocol (for example, if the
parameters are within the preset limits, whether successive reagents
are compatible, etc.). If there are any problems, the screen displays
an appropriate error message.
OK
will not be an option if there are
issues with the protocol.
16. Select the option “Open User” with the appropriate button that is
located next to the log list if you want to open the chamber during
an active protocol only at the “User” Level
Note!
The “Open User” or “Open supervisor” applies to all protocols in each
chamber.
Note!
The device can not detect whether the defined protocol is suitable for the
selected samples.
You must therefore always consult the General Procedures for Tissue
processing and check the compatibility of the individual samples with the
reagents.