Thermo Fisher Scientific
User Guide STP420D/STP420ES
Note!
Paraffin level: You must make sure that the fully melted paraffin is at the
correct level. If needed, add liquid or solid paraffin.
Note!
Paraffin pellets: It takes approximately 4 kg of pellets to fill each container.
4. Check the status window at top right of the monitor. The ‘melting
timer’ indicates that the paraffin is still melting. When this
disappears, the paraffin is ready for operation. When you are sure
that all the paraffin is liquid, you can reset the melting timer. Press
the melting timer icon and
Delete/Turn off
.
5. Now go to Section 2-4-4 Filling the wash bottle
.
Note!
Water level: You can see the filling height of the water inside the bottle in
the oval window on the right side of the instrument. If you have access to the
rear of the instrument, shining a flash light through the rear panel can help
you determine the water level more easily.
1. Open the rear cover above the wash bottle:
2. Remove the screw cap from the wash bottle and insert the funnel
provided into the opening.
Setting up the StP420
Preparing the STP420 and Turning It on
2-4-4 Filling the wash
bottle
Figure 15.
Step 1 for filling the wash bottle.
[1] Rear cover
[2] Screw cap
[2]
[1]