18
TissueWave 2 Operator Guide
The Shandon TissueWave 2 has several features designed to
maximise the efficiency and quality of specimen processing.
These features are as follows:
Ventilation
Vacuum Accumulator
Temperature Probe
Vacuum Chamber
Air agitator
Interior Light
The Shandon TissueWave 2 has a
Vent System Interlock
that
prevents the unit from operating if there is an inadequate air flow
established through the microwave chamber.
This prevents the accumulation of dangerous and toxic fumes.
The vent system on the Shandon TissueWave 2 is designed to be fed
to a
Powered Exhaust System
.
A
4” (100 mm) Chamber Outlet
is attached to the back of the unit
for easy exhaust hook-up.
An
Aluminium 4” (100 mm) Diameter Hose
is included with the
unit for ventilation.
Note:
It is recommended that aluminium hosing be used since plastic hoses have a
tendency to collapse at bends.
•
•
•
•
•
•