26
TissueWave 2 Operator Guide
Note:
It is possible to engage the vacuum prior to the protocol step
beginning, using the Vacuum Control Key on the Keypad.
This will ensure that the lid is properly seated and that a vacuum
has been established before the protocol step begins.
When fully engaged, the
Vacuum Gauge
should read
approximately
-15 inHg
, although this may be affected by
load and run volume.
Connections for Processing Without Vacuum:
Mount
Temperature Probe
so that the tip is at about the
Half-Way Point
of the Depth of Fluid, and as close to the
centre as possible and not in contact with any solid material.
Connect the
Air Agitator Tube
to the coupling adjacent to the
Agitator Valve
on the
Microwave Chamber
roof.
Immerse the open end of the
Air Agitator Tube
in the
Processing Fluid
.
The
Vacuum
and
Exhaust Ports
should be left
Open
.
•
•
•
•
•
Exhaust
Port
Vacuum
Port
Air
Agitator
Temperature
Probe
Vacuum
Chamber
Base
Agitator
Valve
Exhaust
Port
Vacuum
Port
Air
Agitator
Temperature
Probe
Vacuum
Chamber
Base
Agitator
Valve