44
TissueWave 2 Operator Guide
There are also some faults which may occur, for which no
Warning
Screen
is provided.
The following table lists those most commonly encountered:
Other Faults
Problem
•
Check that the Tubing is properly connected to the
Chamber Lid
•
Check that the Chamber Lid is securely fitted and makes
a good contact at all points
•
Check the Chamber Lid mating surfaces for damage
•
Check the Bushing around the Temperature Probe
•
Check that the Vacuum Control Knob is set correctly
•
Check that the Vacuum Gauge is working correctly
•
Check that the seal between the Vacuum Chamber and
its Lid is clear of debris.
•
If problems persist call ThermoFisher Scientific
Technical Support
•
Check that the Temperature probe is correctly
positioned
•
Check the Power Settings
•
Check that the Load is correct for the Power Setting
•
Check that the Preset Temperature has not been
changed
•
Use marble chips if necessary
•
If using paraffin (wax), ensure the temperature of the
molten paraffin is close to the Preset Temperature
before processing
•
If problems persist call ThermoFisher Scientific
Technical Support
Unable to draw a vacuum on the vacuum
chamber
Remedy
Unable to achieve set temperature