2-2
|
Precision
Thermo Scientific
Chapter 2
|
General Information
General Purpose Water Bath Specifications
GP 02
GP 2S
GP 05
GP 10
Reservoir Fluid Control
Temperature
°C
°F
Ambient to 90
Ambient to 194
Ambient to 100
Ambient to 212
Set Point Temperature
Range
°C
°F
+5 to 100
+9 to 212
Ambient Temperature
°C
Range
°F
+15 to 45
+59 to 113
Stability
@
37°C
±0.1
Uniformity
@
37°C
±0.2
Heating Output
watts
200
300
300
800
Bath Volume
liters
2
2
5
10
Overall Bath Dimensions
(L x W x H) cm
inches
23.0 x 19.9 x 23.3
9.1 x 7.8 x 9.2
24.6 x 35.5 x 23.2
9.7 x 14.0 x 9.1
24.6 x 35.5 x 23.2
9.7 x 14.0 x 9.1
39.3 x 38.3 x 23.3
15.5 x 15.1 x 9.2
Bath Work Area
Dimensions
(L x W x H) cm
inches
13.8 x 15.5 x 15.0
5.4 x 6.1 x 5.9
15.3 x 30.0 x 6.5
6.0 x 11.8 x 2.6
15.4 x 30.0 x 15.0
6.1 x 11.8 x 5.9
30.1 x 33.0 x 15.0
11.9 x 13.0 x 5.9
Approximate Weight
kg
lb
3.5
7
4
9
4.5
10
7.5
16
Electrical
Requirements
(VAC/Hz)
(Voltage ±10%)
100-115/50-60
or
200-230/50-60
Compliance
CE RoHS
UL WEEE
Maximum Relative
Humidity
(Non Condensing)
80% for temperatures up to 31°C (88°F)
decreasing linearly to 50% relative humidity at 40°C (104°F)
Operating Altitude
meters
feet
Sea Level to 2000
Sea Level to 6560
Overvoltage Category
II
Pollution Degree
2
Storage Temperature
°C
Range
°F
-25 to +60
-13 to +140
Low-end reservoir fluid control temperatures require supplemental cooling.
Specifications obtained at sea level using water.
Thermal beads may be used instead of water but they will degrade the bath's uniformity and stability.
Bath depth includes bezel, height does not include lid.
Thermo Fisher Scientific takes no responsibility for damages caused by the selection of an unapproved fluids.
Thermo Fisher Scientific reserves the right to change specifications without notice.
Summary of Contents for Precision GP 02
Page 24: ...DANGER WARNING CAUTION I II III NEC 100 80 40 C 50 50 O 60 C BG...
Page 27: ...DANGER WARNING CAUTION I II III 100 80 C 40 C 50 50 O off 60 C EL...
Page 45: ...2 4 Precision Thermo Scientific Chapter 2 General Information...
Page 61: ...4 8 Precision Thermo Scientific Chapter 4 Operation...
Page 63: ...5 2 Precision Thermo Scientific Chapter 5 Preventive Maintenance...
Page 66: ......