Chapter 4
pH, pH/ISE and pH/LogR Measurement Modules
50
Orion Versa Star Pro Meter User Manual
Thermo Scientific
mV, Relative mV and ORP Mode Menus
Main Menu
Settings
Details
Electrode SI No
----
Use the alphanumeric keypad to enter the electrode
serial number or clear all to shown no value (----)
Sample ID
Off, Manual, Auto Increment
Use the popup keypad screen to enter an
alphanumeric manual ID or a numeric value as the
starting value for auto incremental ID
Stability
Smart, Fast, Medium, Slow Set when a measurement is recognized as stable
Averaging
Off, Automatic Smart
Set averaging for faster measurement stability
Read Type
Auto-Read, Timed,
Single-Shot, Continuous
For Timed and Single-Shot, set the time in hours,
minutes, seconds (HH:MM:SS) using the arrow keys
Alarm*
Limit Alarm
Off, On
Set high and low limit values to activate alarm
Cal Due Alarm Off, On
Set a calibration interval (hours) to activate alarm
Set Point Alarm Off, On
Set a base and offset value to activate alarm
* The mV mode menu will not have the Cal Due Alarm option.
Detailed information on these features is available in the
Appendix
section of this user manual.