Chapter 8
Data Transfer and Software Updates
Thermo Scientific
Orion Versa Star Pro Meter User Manual
105
To Get
Continuous
Measurements:
1.
SETCHANNEL X <CR>
a.
X = Channel number
b.
For multiple channels, enter channel numbers without commas
c.
Example: SETCHANNEL 1 <CR>
2.
SETREADTYPE X, 4 <CR>
a.
X = Channel number
b.
Y = Read Type (4 = Continuous)
c.
Z = leave blank
d.
Example: SETREADTYPE 1, 4 <CR>
3.
GETMEAS <CR>
a.
Prints the displayed channel measurement immediately and
continuously until the STOP <CR> command is given
b.
Example: GETMEAS <CR>
SYSTEM
SYSTEM <CR>
Prints the Meter Type, Serial Number, Software Version
Example: Versa Star Pro, V11702, 11.20
SETRTC
SETRTC YY MM DD HH MM SS <CR>
Set the date and time (in 24-hour format) for the meter
Example: SETRTC 18 03 29 15 00 00
GETCHCONFIG
GETCHCONFIG <CR>
Prints the meter channel configuration with mode details
GETMODE
GETMODE CHANNEL <CR>
Print the channel measurement mode
Example: GETMODE 1
ESC
ESC <CR>
Stops the command being executed
STOP
STOP <CR>
Command stops continuous measurement from meter
Meter Measurement Data Example
Mode
Format Example
pH
Meter Type, Meter Serial Number, Meter Software Revision, Date, Time, Channel
Number, Electrode ID, Sample ID, User ID, Module Serial Number, Mode, pH Value, pH
Unit, mV Value, mV Unit, Temperature Value, Temperature Unit, Slope Value, Slop Unit,
Method Number
Example: Versa Star Pro, V11702, 11.20, 29-03-18, 3:50:53 PM, Ch-1, SENSOR-01, 1,
ADMIN, VA12345, pH, 7.000, pH, 0.0, mV, 25.0, C (MAN), 100.0, %, M101 <CR>