
Chapter 8
Data Transfer and Software Updates
Thermo Scientific
Orion Versa Star Pro Meter User Manual
103
GETDATALOG
GETDATALOG <CR>
Prints all the logged measurement data
If no data is available, will return “>” to receive next command
GETDATALOG X, Y
X = First data log number
Y = Last data log number
Prints the logged measurement data in a specified range
If no data is available, will return “>” to receive next command
Example: GETDATALOG 10, 100 <CR>
GETCALLOG
GETCALLOG X <CR>
Prints all the logged calibration data
If no calibration is available, will return “>” to receive next command
X = Select calibration data for specific mode as a number:
PH = 1
COND = 6
RMV = 3
DO = 10
ORP = 4
RDO = 12
ISE = 5
INCREMENTAL = 21
Example: GETCALLOG 1 <CR>
SETCHANNEL
SETCHANNEL X <CR>
X = Channel number
For multiple channels, enter channel numbers without commas
Updates the displayed channel(s) on the meter
Example: SETCHANNEL 1 <CR>
SETMODE
SETMODE X Y <CR>
Changes the measurement mode of the entered channel
X = Channel number
Y = Measurement mode
PH = 1
COND = 6
DO, % Saturation = 10
MV = 2
TDS = 7
DO, mg/L = 11
RMV = 3
SALINITY = 8
ORP = 4
RESISTIVITY = 9
ISE = 5
Example: SETMODE 1 1 <CR>