Electrode Diagnostics Log
Electrode
Model
Electrode
Serial #
Date
Temp. Resistance
Change in
Resistance
Notes
8102BNWP
(example)
PV1-27726 1/6/14 24.5°C 340 MΩ
---
Initial Check
8102BNWP
(example)
PV1-27726 1/13/14 24.0°C 20 MΩ
320 MΩ
Large resistance decrease,
indicates electrode failure –
crack in pH bulb observed