Chapter 3
Meter Setup Menus
Thermo Scientific
Orion Versa Star Pro Meter User Manual
31
Refer to
Chapter 4 pH, pH/ISE and pH/LogR Measurement Modules
for detailed instructions on
performing a LogR temperature calibration with a glass-bulb pH electrode when using the Versa
Star Pro meter with pH/LogR measurement module.
Manual Temperature Entry
Enter a manual temperature value when MAN is selected in the Temperature Input menu.
1.
In the measurement mode, press the setup key.
2.
Press the
,
,
or
key to highlight Channel 1, Channel 2, Channel 3 or Channel 4
and press the f3 (Select) key.
3.
Press the
or
key to highlight Temperature and press the f3 (Select) key.
4.
Press the
or
key to highlight Temperature Cal and press the f3 (Select) key.
5.
Press the f3 (Clear) key, input the temperature value using the numeric keypad and then
press the f2 (Enter) key to save the value.
6.
Press the measure (esc) key to return to the measurement mode.