2
Install Multiskan Sky
Connect the mains supply cable
Thermo Scientific
Thermo Scientific Multiskan Sky User Manual
7
6. Attach the transport lock on the back of the instrument with the red fixing screw and the locking
piece that is on the back of the instrument.
Connect the mains supply cable
1. Connect the power supply cable to the power supply connector and plug in the instrument.
2. Connect the power supply to a correctly installed line power outlet with a grounded conductor.
Connect the instrument to a computer
If you are using a computer with the instrument, connect the communication cable to the USB port
marked USB/PC.
Install SkanIt Software
For installation instructions, refer to
“Install SkanIt Software”
on
page 9
.
When the instrument is connected to SkanIt Software, a
Take control
icon shows on the display. Tap
the icon to disconnect the computer remote control.
Operational check
When you switch the instrument on the instrument proceeds to perform the self diagnostics. No error
messages should appear on the display.
CAUTION
Do not use other power supply cables than the power supply cable delivered with the
instrument. Use the Thermo Scientific power supply cable designed for your region.
CAUTION
Do not operate your instrument from a power outlet that has no ground connection.
Note
Connecting the instrument to SkanIt Software automatically updates the instrument date
and time according to the PC clock.