28
Calibration
• Pause consistently.
It takes a moment for the liquid in the tip to finish
moving. Pause with the tip in the liquid for about one second after
aspirating the sample to allow the liquid to finish moving into the pipette
tip after the pipette stops. Otherwise, the volume will be lower than it
should.
• Aspirate slow, dispense fast.
As a general rule, pipette liquids,
especially viscous liquids, at a slower speed to ensure accurate
pipetting. Dispense aqueous solutions at faster speeds to expel liquid
more accurately.
Note:
Viscous liquids should be dispensed slowly.
• Tip touch.
During the dispensing of a liquid, touch the tip against the
side of the receiving container to prevent any liquid from remaining on
the tip.
• Discard final dispense.
When accurate dispensing of several
increments of liquid are needed, discard the last increment. The
cumulative percentage of error from previous increments carries to the
last volume. (i.e. The pipette aspirates 100 µl and dispenses 20 µl five
times, and if each of the first four dispenses were off by +1.0 µl, then the
last dispense is missing 4 µl before you even dispense it).
Summary of Contents for Matrix MultiChannel Equalizer Pipette
Page 5: ...IV ...
Page 57: ...52 Index ...