26
Calibration
4. Press the [Vol] key to display “CALC”
(pressing the [Vol] key again will
return you to the “EASY” mode)
.
5. Press the [Enter] key to display the calibration point.
6. Press the [Vol] key to enter the Display Volume
(Dv)
calculated above in
step 2.
7. Press the [Enter] key to set the calibration point. The display shows “CAL”
then “SET” to confirm the new calibration point, and returns to the current
pipetting program.
Example
: Calibrate a 1250 µl 8-Channel pipette.
Water temperature = 24 °C
Programmed Volume (Vp) = 1000 µl
Average of Actual Volume = 990 µl
Corrected Actual Volume (Va) = 990 x 1.004 = 994 µl
Calibration Point (Cp) = 1000
D
V
= x Cp
V
p
Va
D
V
= x
1000
1000
µl
994
µl
D
V
= 1006
D.2.4 CALVIEW
CALVIEW displays the pipetting stroke
(number of steps)
used to reach
the calibration point. This value allows you to confirm the calibration during
pipetting and after calibration. To activate CALVIEW, enter the CAL mode
using the Special Function key. When the calibration point is displayed,
press the [Fill] key. The number of steps is displayed. Press any key to exit
CALVIEW.