7.
Using the tool provided on the inside of the back panel (Figure A), secure the
light cube by tightening the two screws on the end of the cube (red arrows in
8.
To remove a light cube, unscrew both screws that secure it to the instrument.
9.
Thread the light cube removal tool into the central hole in the cube (white arrows
in Figure B) and gently pull the light cube out of the device.
Note:
Always store the cube removal tool in the back panel for easy access.
10.
Install the back panel and secure it in its place with both ¼-turn fasteners.
11.
Plug the power cord back into the Countess
™
II FL Automated Cell Counter.
12.
Turn off the Countess
™
II FL Automated Cell Counter by flipping the power
switch on the back of the instrument to the ON position.
Chapter 8
Instrument settings
Change light cube
8
Countess
™
II FL automated cell counter User Guide
55