HistoStar Operator Guide A81010100 Issue 2
78
Wax Tank and Dispense
Problem
Possible Cause
Remedy
Wax will not dispense
Insufficient level of wax
in wax tank
Add wax to wax tank
Wax has not yet melted Allow time for wax to melt
Ensure temperature setting is high enough to melt the wax
being used
Compare the set point temperature to the actual temperature -
if the set point temperature is not achieved after a suitable
length of time:
•
Switch off the embedding module
•
Wait for 20 minutes and then switch the embedding
module back on
If problem persists call Thermo Fisher Scientific Service
Department
Wax dispense flow
control is closed
Open the flow control by turning the knob counter-clockwise
Instrument
malfunction
Switch the embedding module off for 30 seconds and then
back on again to reset the control system
Perform Heater System Check
Call Thermo Fisher Scientific Service Department
Wax will not dispense
when using footswitch
Disconnected or poorly
fitted air tube
Check that air tube is fully inserted into footswitch connector
port - see Connections
Ruptured tubing or
footswitch bellows
Replace footswitch
Erratic temperature
control
Instrument
malfunction
Switch the embedding module off for 30 seconds and then
back on again to reset the control system
Compare the set point temperature to the actual temperature -
if the set point temperature is not achieved after a suitable
length of time:
•
Switch off the embedding module
•
Wait for 20 minutes and then switch the embedding
module back on
If problem persists call Thermo Fisher Scientific Service
Department
Summary of Contents for HistoStar
Page 1: ...Thermo Scientific HistoStar Operator Guide A81010100 Issue 2 ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 12: ...HistoStar Operator Guide A81010100 Issue 2 4 Product Return Safety Declaration ...
Page 14: ......
Page 34: ......
Page 98: ......
Page 118: ......
Page 122: ...HistoStar Operator Guide A81010100 Issue 2 114 Appendix C Screen maps ...
Page 124: ......
Page 128: ...120 ...