HistoStar Operator Guide A81010100 Issue 2
106
Draining the Wax Tank
To drain the wax tank:
•
Ensure the wax in the wax tank is molten.
Notes
It is advisable to ensure that the hot-spot and the tissue storage tank are switched on when draining as this will
ensure that the wax remains molten until it has exited the internal pipework.
It is recommended that the wax tank should be no more than half full before draining. If the wax tank is too full,
manually remove wax from the wax tank by opening the wax tank cover and bailing out the wax using a suitable
vessel.
•
Completely remove the waste wax tray from the unit.