HistoStar Operator Guide A81010100 Issue 2
24
Using the Para Trimmer
The Para Trimmer surface on the Embedding Module can be used to remove excess wax from an embedded
cassette.
To clean wax from an embedded cassette:
•
Ensure the hot spot is at its preset operating temperature - the Para Trimmer temperature is determined by
the hot spot settings.
•
Hold an embedded cassette between thumb and forefinger, and place one of the edges against the Para
Trimmer surface.
•
As the wax begins to melt, drag the edge of the cassette across the serrated face of the Para Trimmer to
scrape off the excess wax.
The surface of the Para Trimmer is hot and care should be taken
to avoid skin contact during use.
The Para Trimmer edges are sharp. Incorrect use can cause
injury and / or specimen damage.
•
Repeat the above procedure for all edges until the cassette is clean.
Notes
The Para Trimmer is safe for cleaning the printed surface of a cassette - print will not be removed during correct
use.
The back end of some cassettes have a notch to allow a lid to lock into; this can be cleaned by running the notched
edge along the length of one of the serrations.
The Para Trimmer can be cleaned during operation by wiping the surface with a soft, absorbent towel.
Summary of Contents for HistoStar
Page 1: ...Thermo Scientific HistoStar Operator Guide A81010100 Issue 2 ...
Page 2: ......
Page 4: ......
Page 6: ......
Page 12: ...HistoStar Operator Guide A81010100 Issue 2 4 Product Return Safety Declaration ...
Page 14: ......
Page 34: ......
Page 98: ......
Page 118: ......
Page 122: ...HistoStar Operator Guide A81010100 Issue 2 114 Appendix C Screen maps ...
Page 124: ......
Page 128: ...120 ...