A1.5 Log
A1.5.1 Logging
Notes
•
The Log Index is a list of pages of the records or “Logs” in the Sarasota 200
•
There are 8 pages available. Logging will automatically continue on the following pages
until the whole log memory is full at which point it will stop or wrap round. See A1.5.1.1
•
.As an example, 12 variables at 15 minute logging intervals will fill the memory in 210
days.
•
The “Log Header” shows the details of the highlighted Log page
•
The Log Size indicates the number of record samples and the proportion of the page
occupied by the highlighted Log.
•
“Refresh Index” causes the latest status of the selected Log to be displayed. This will be
changing if the Sarasota 200 is operating.
•
Clicking “Collect Log” brings up a “Save As” screen to allow the Log to be saved to the
computer memory as a comma separated variable (.txt) file.
•
The saved file may be imported into a spread sheet.
•
“Erase Log” clears all the Logs.
•
“Config Log” allows up to 12 parameters to be selected for logging via the “Log
Configuration” screen A1.5.1.1.