Sarasota 200 Ultrasonic Multipath
Flowmeter
Appendix 1 GAFA PC Software Page A1-15
Thermo Fisher Scientific
APPENDIX 1 GAFA PC SOFTWARE
A1.4.2.4 (Channel Configuration) – Profile
Notes
•
This table defines the cross section shape of the channel.
•
It is entered as a series of widths against heights above datum.
•
The Sarasota 200 will interpolate between the entered values.
•
For channels with straight sides (even if not vertical), it is only necessary to enter 2
points, the width at the bottom and at the top.
•
It is important to enter a width for a height greater than or equal to the maximum water
level because the cross section area of the channel will only be calculated up to the
highest defined width.
•
Click on “Write To Gauge” to confirm selections to the Sarasota 200 or “Cancel” to
abandon.